HASO™3 WSR (Wide Spectral Range) wavefront sensors

HASO3 WSR (Wide Spectral Range) is the latest addition to our award winning HASO product family.  Available in two configurations, 32 x 40 and 58 x 74 microlenses, they combine the standard functionalities that customers have come to rely on, including absolute measurement, wide dynamic range, unequalled accuracy and insensitivity to vibration, with the capacity to measurement wavefronts across an exceptionally wide spectral band, even with monochromatic light.

Our proprietary Dynamic Spot Tracking™ and Auto Spot Finder™ features combined with the exceptional optical quality of our refractive microlenses in our patented Rotated Square™ configuration provide superior accuracy and eliminate crossing spots, without compromising dynamic range.  Even more, Imagine Optic’s proprietary calibration technology enables you to obtain absolute measurements every time, without the need for a reference beam, and our patented sensor design allows you to measure phase and intensity independently and in real-time.  When coupled with our HASO v3 software package, you can perform both zonal and modal wavefront reconstruction; calculate the PSF, MTF and Strehl ratio; visualize the spot diagram; and obtain the M² parameter. 

HASO3 WSR is ideally suited for applications including laser beam characterization (particularly adapted for frequency doubling), optical system metrology over a large spectral band, measuring the effects of chromatic aberrations and aligning laser optics.

If you would like more information on our products, please call +33 (0)1 64 86 15 60 or e-mail us at .

 

  Applications
    Pure science
 
  Documentation
    Click the PDF icon to download the product brochure.

  HASO3 WSR-32 HASO3 WSR-58
Aperture dimension 4.9 x 6.1mm² 8.9 x 11.3
Refractive microlens technology Rotated square
Sub-apertures dedicated for analysis 32 x 40 58 x 74
Tilt dynamic range >±3° (520λ)
Focus dynamic range - minimum local
radius of curvature
20 mm 15mm
Focus dynamic range - maximum n.a. > 0.1
Repeatability < λ/200

Wavefront measurement accuracy
in relative mode (rms) (1)

λ/150
Wavefront measurement accuracy
in absolute mode (rms) (2)
λ/100
Tilt measurement accuracy (rms) 3 µrad 1 µrad
Focus measurement accuracy (rms) 10-3 m-1 5.10-4 m-1
Spatial resolution

~160 µm

Working wavelength range 350 - 1100 nm
Calibrated wavelength range 400 - 800 nm or 532 - 1064 nm
Interface FireWire FireWire and CameraLink
Maximum acquisition frequency 50 Hz 7Hz FireWire, 16Hz
Processing frequency (CPU 3Ghz, 512 Mb RAM) 20 Hz 10 Hz
Working temperature 15 - 30° C
Dimension / weight 75 x 62 x 68 mm / 510 g 70 x 55 x 60 mm / 258 g (camera)
1.1 Kg (controller)
Power supply 12 V / 6 W
1) Difference between the real wavefront and a reference wavefront obtained in similar conditions (10 λ of shift maximum).
2) Wavefront as seen by the wavefront sensor. Performance kept on the whole spectral range.

 

© 2007 Imagine Optic. All Rights Reserved. HASO, Mirao and other product names mentioned on this site are or may be trademarks